Product Specifications
-
Cleaning method:
Soaking, jet cleaning, ultrasonic cleaning
-
pH:
Weak alkaline
-
Dilution rate:
- Soaking 1.0% (100mL per 10L of warm water)
- Jet cleaning 0.5% (50mL per 10L of water)
- Ultrasonic cleaning 1.0% (100mL per 10L of water) -
Foaming:
Low foaming
-
Rinse:
Necessary
-
Compatibility:
Stainless steel, copper, brass and aluminum
Frequently Asked Questions
-
How to use Power Quick Multi-Purpose Enzyme Cleaner
Manual Cleaning
- Prepare a dilution with warm water (around 40°C).
- The recommended concentration for use is 1.0% (100mL/10L of water).
- Adjust between 0.5% and 1.0% (50 to 100mL/10L of water) depending on the degree of soil.
- Immerse the used medical instruments in the dilution for at least 10 minutes.
- Rinse completely in running water.
Ultrasonic Cleaning
- Prepare a dilution with water in an ultrasonic washer basin.
- The recommended concentration for use is 1.0% (100mL/10L of water).
- Adjust between 0.5% and 1.0% (50 to 100mL/10L of water) depending on the degree of soil.
- Put the used instruments into the basin, then wash for 10 to 15 minutes according to the manufacturer's instructions for the washer.
- Rinse completely with running water.
Jet Cleaning
- Pump a dilution into a jet washer.
- The recommended concentration for use is 0.5% (50mL/10L of water).
- Adjust between 0.2% and 0.5% (20 to 50mL/10L of water) according to the degree of soil.
- Put the used instruments in the jet washer.
- Operate the jet washer according to the manufacturer's instructions.
- Prepare a dilution with warm water (around 40°C).