Power Quick Multitype Enzyme Cleaner

Ultrasonic

Dilution

Formulated with 4 different types of enzymes which help effectively dissolve and remove a variety of soils, and sophorolipid, a natural biosurfactant that helps prevent redeposition of soils.

  • Slightly alkaline enzyme cleaner for manual cleaning of medical instruments. Formulated with 4 different enzymes to effectively remove a variety of soiling in addition to sophorolipids*, a natural biosurfactant to assist in cleaning.

  • It can be used for manual cleaning or in ultrasonic machines due to the low foaming formulation. Sophorolipids rinses cleanly away and prevents recontamination.

  • Non-corrosive to many metals.

  • The 4 enzymes are carefully formulated to maximize the stability under alkaline conditions while maintaining good anti-corrosive effect and detergency for long periods of time.

  • Color formulation can visually indicate when solution has been added to water.

*

Sophorolipids are a natural biosurfactant produced by fermentation of the yeast Starmerella bombicola with palm oil and glucose. It is highly biodegradable and has strong detergency.

Product Specifications

  • Cleaning method:

    Soaking, ultrasonic cleaning

  • pH:

    Weak alkaline

  • Dilution rate:

    1.0% (100mL per 10L of warm water)

  • Foaming:

    Low foaming

  • Rinse:

    Necessary

  • Compatibility:

    Stainless steel, copper, brass and aluminum


Frequently Asked Questions

  • How to use:

    Manual Cleaning
    1. Prepare a dilution with warm water (around 40°C).
    - Recommended concentration for use is 1.0% (100mL/10L of water).
    - Adjust between 0.5% and 2.0% (50 to 200mL/10L of water) depending on the degree of soil.
    2. Immerse used medical instruments in the dilution for at least 10 minutes.
    3. Rinse completely in running water.

    Ultrasonic Cleaning
    1. Prepare a dilution with water in an ultrasonic washer basin.
    - Recommended concentration for use is 1.0% (100mL/10L of water).
    - Adjust between 0.5% and 2.0% (50 to 200mL/10L of water) depending on the degree of soil.
    2. Put used instruments into the basin, then wash for 10 to 15 minutes according to the manufacturer's instructions for the washer.
    3. Rinse completely with running water.